FIGURE 18-16 Right, Quadrant-sharing detector, in which each scintillator block straddles the corners of four photomultiplier tubes (PMTs). This allows larger PMTs to be used as compared to a standard block detector (left). For large panels, this leads to almost a fourfold reduction in the number of PMTs required to read out a given number of scintillator elements.